gms | German Medical Science

68th Annual Meeting of the German Society of Neurosurgery (DGNC)
7th Joint Meeting with the British Neurosurgical Society (SBNS)

German Society of Neurosurgery (DGNC)

14 - 17 May 2017, Magdeburg

Assessment of rechagring, satisfaction, confidence and adverse Events of rechargeable internal pulse Generators as Initial neurostimulators in patients with deep brain Stimulation for movement disorders

Meeting Abstract

  • Martin Jakobs - Neurochirurgische Klinik, Universitätsklinikum Heidelberg, Heidelberg, Deutschland
  • Manja Kloß - Neurologische Klinik, Universitätsklinikum Heidelberg, Heidelberg, Deutschland
  • Andreas Unterberg - Neurochirurgische Klinik, Universitätsklinikum Heidelberg, Heidelberg, Deutschland
  • Karl Kiening - Neurochirurgische Klinik, Universitätsklinikum Heidelberg, Heidelberg, Deutschland

Deutsche Gesellschaft für Neurochirurgie. Society of British Neurological Surgeons. 68. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie (DGNC), 7. Joint Meeting mit der Society of British Neurological Surgeons (SBNS). Magdeburg, 14.-17.05.2017. Düsseldorf: German Medical Science GMS Publishing House; 2017. DocMO.16.02

doi: 10.3205/17dgnc092, urn:nbn:de:0183-17dgnc0925

Published: June 9, 2017

© 2017 Jakobs et al.
This is an Open Access article distributed under the terms of the Creative Commons Attribution 4.0 License. See license information at http://creativecommons.org/licenses/by/4.0/.


Outline

Text

Objective: Rechargeable internal pulse generators (r-IPGs) for deep brain stimulation (DBS) promise a longer battery life and cost effectiveness compared to non-rechargeable IPGs. However, patients need to learn to check the battery capacity and perform the recharging process to ensure continuous therapy.

Methods: n=35 consecutive adult patients with movement disorders that underwent DBS electrode placement with implantation of a r-IPG were assessed with a standardized questionnaire. They were asked to report on their recharging routine, user confidence, satisfaction and adverse events. Patients were asked to assess the level of difficulty of the individual steps and the overall recharging process on an ordinal scale awarding 1 to 5 points.

Results: 89% (n=31) patients responded and were available for data analysis. n=21 patients received DBS for Parkinson's Disease, n=8 for essential tremor and n=2 for dystonia at a mean age of 63.3 years. The mean follow up was 21.2 months. n=7 patients have partners or nursing services check and recharge the IPG. The recharging takes an average of 57.6 minutes. 90,3% felt confident using their IPG after a mean of 2.1 weeks and 1.6 training sessions. 97% of patients prefer their r-IPG over a conventional one. n=3 patients experienced inability to recharge their IPG at some point. One patient experienced battery depletion and interruption of stimulation because of inability to recharge. The overall recharging process as well as each individual step was rated as „easy“ at a median score of 4.0 out of 5 points. Old age was not associated with more adverse events or a lower rating for the recharging process.

Conclusion: Choosing a r-IPG during initial DBS surgery is safe and associated with a low number of adverse events even in older patients. Handling and recharging the IPG is considered as "easy" by the vast majority of patients. Most of the patients undergoing DBS for movement disorders will profit from the advantages of r-IPGs.