gms | German Medical Science

66th Annual Meeting of the German Society of Neurosurgery (DGNC)
Friendship Meeting with the Italian Society of Neurosurgery (SINch)

German Society of Neurosurgery (DGNC)

7 - 10 June 2015, Karlsruhe

Investigating intracranial pressure of patients with normal pressure hydrocephalus using the Aesculap-Miethke SensorReservoir

Meeting Abstract

  • Philipp Ertl - Klinik für Neurochirurgie, Medizinische Hochschule Hannover, Hannover, Deutschland
  • Elvis J. Hermann - Klinik für Neurochirurgie, Medizinische Hochschule Hannover, Hannover, Deutschland
  • Hans Heissler - Klinik für Neurochirurgie, Medizinische Hochschule Hannover, Hannover, Deutschland
  • Joachim K. Krauss - Klinik für Neurochirurgie, Medizinische Hochschule Hannover, Hannover, Deutschland

Deutsche Gesellschaft für Neurochirurgie. 66. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie (DGNC). Karlsruhe, 07.-10.06.2015. Düsseldorf: German Medical Science GMS Publishing House; 2015. DocP 170

doi: 10.3205/15dgnc568, urn:nbn:de:0183-15dgnc5689

Published: June 2, 2015

© 2015 Ertl et al.
This is an Open Access article distributed under the terms of the Creative Commons Attribution 4.0 License. See license information at http://creativecommons.org/licenses/by/4.0/.


Outline

Text

Objective: The dynamics of intracranial pressure (ICP) after shunt surgery in patients with normal pressure hydrocephalus is poorly known. Temporarily implanted intraparenchymal pressure probes have shown to be reliable for precise ICP-monitoring, however, a disadvantage of this method is that the ICP probe has to be explanted after a period of time requiring additional surgery. We present two patients with normal pressure hydrocephalus with an integrated ICP measuring device in the shunt system which allows for long-term postoperative telemetric monitoring of ICP.

Method: Two patients (1 man, 66-year-old, 1 woman, 78-year-old) with normal pressure hydrocephalus underwent shunt surgery using the Aesculap-Miethke SensorReservoir as an ICP measuring device integrated in the shunt system. On the first postoperative day and during the follow-up examination 3 months later, several measurements of ICP were performed using the handheld device to read the SensorReservoirs data. Postural changes such as lying, sitting and standing with different head postures were assessed according to an experimental protocol in a randomized order during which the ICP was measured.

Results: There were clear ICP responses to postural changes, in line with physiologically expected values. Since the highest measurement rate of the Aesculap-Miethke SensorReservoir is 1 Hz, however, the collected data has to be considered as an approximation for actual ICP dynamics.

Conclusions: The Aesculap-Miethke SensorReservoir is an easy-to-use tool to reliably measure intracranial pressure in patients with a shunt system. The sample rate and handling of data acquisition may be further developed. As for the time being, it is difficult to draw therapeutic consequences from interpreting the acquired data.