gms | German Medical Science

30. Kongress der Deutschsprachigen Gesellschaft für Intraokularlinsen-Implantation, Interventionelle und Refraktive Chirurgie (DGII)

Deutschsprachige Gesellschaft für Intraokularlinsen-Implantation, Interventionelle und Refraktive Chirurgie (DGII)

11.02. - 13.02.2016, Mannheim

Relation of pupil size and cataract surgery using PupilX

Meeting Abstract

  • Annekatrin Rickmann - Sulzbach/Saar
  • M. Waizel - Sulzbach/Saar
  • S. Kazerounian - Sulzbach/Saar
  • P. Szurman - Sulzbach/Saar; Tübingen
  • H. Wilhelm - Tübingen
  • K.T. Boden - Sulzbach/Saar

Deutschsprachige Gesellschaft für Intraokularlinsen-Implantation, Interventionelle und Refraktive Chirurgie. 30. Kongress der Deutschsprachigen Gesellschaft für Intraokularlinsen-Implantation, Interventionelle und Refraktive Chirurgie (DGII). Mannheim, 11.-13.02.2016. Düsseldorf: German Medical Science GMS Publishing House; 2016. Doc16dgii047

doi: 10.3205/16dgii047, urn:nbn:de:0183-16dgii0471

Published: March 1, 2016

© 2016 Rickmann et al.
This is an Open Access article distributed under the terms of the Creative Commons Attribution 4.0 License. See license information at http://creativecommons.org/licenses/by/4.0/.


Outline

Text

Background: The aim of this study was to evaluate the influence of cataract surgery on pupil size at well defined measurement conditions with a novel pupillometer (PupilX).

Methods: Pupil size of healthy study participants was measured with infrared-video PupilX pupillometer (MEye Tech GmbH) at 5 different illumination levels (0, 0.5, 4, 32 and 250 lux) before and after cataract surgery. Measurements were performed by the same person during a measurement period of 3 seconds with 90 images and an absolute linear camera resolution of approximately 20 pixel per mm.

Results: This prospective study analysed the pupil size of 35 patients (mean age 73.2 ± 8.6 years, range 50-87 years). One week after cataract surgery mean pupil diameter for all patients decreased ignificantly by -13.1 % at 0 Lux, -13.6% at 0.5 Lux, -8.4% at 4 Lux, -7.3% at 32 Lux and -5.1% at 250 Lux. Beside the gradually decreasing pupil size with an increasing illumination level there is a significant decreased pupil size after cataract surgery at all illumination levels.

Conclusion: In conclusion, this study showed a significant decrease of pupil size after cataract surgery in a normal population under scotopic, mesopic and photopic static illuminance conditions. The precision of measurements and variability of illuminance conditions of PupilX (MEye Tech GmbH) allows the possibility of investigations of pupil size in relation to refractive surgery.