LASIK-induced changes of corneal polarization axis and magnitude can be compensated by scanning-laser polarimetry with variable cornea compensator

Meeting Abstract

  • corresponding author N. Niederdräing - University Eye Hospital, Essen
  • C. Heinz - University Eye Hospital, Essen
  • J. M. Selbach - University Eye Hospital, Essen
  • T. Laube - University Eye Hospital, Essen
  • K.-P. Steuhl - University Eye Hospital, Essen
  • S. Kremmer - University Eye Hospital, Essen

Evidenzbasierte Medizin - Anspruch und Wirklichkeit. 102. Jahrestagung der Deutschen Ophthalmologischen Gesellschaft. Berlin, 23.-26.09.2004. Düsseldorf, Köln: German Medical Science; 2004. Doc04dogP 102

Die elektronische Version dieses Artikels ist vollständig und ist verfügbar unter: http://www.egms.de/de/meetings/dog2004/04dog593.shtml

Veröffentlicht: 22. September 2004

© 2004 Niederdräing et al.
Dieser Artikel ist ein Open Access-Artikel und steht unter den Creative Commons Lizenzbedingungen (http://creativecommons.org/licenses/by-nc-nd/3.0/deed.de). Er darf vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden, vorausgesetzt dass Autor und Quelle genannt werden.


Gliederung

Text

Objective

Corneal birefringent properties can influence scanning laser polarimetry (SLP) with fixed compensator. Recently, a new variable cornea compensator was introduced. We investigated wether the new device allows to measure and to compensate changes of corneal polarization axis and magnitude before and after LASIK. The results are compared to scanning laser polarimetry measurements with fixed cornea compensator.

Methods

Twenty-six eyes of 16 patients were enrolled in this prospective study. Before myopic LASIK surgery all patients underwent standard ophthalmological examinations, corneal topography, corneal pachymetry, and assessment of pupil diameter. Scanning laser polarimetry was performed with a GDxVCC (LDT, San Diego, USA, software version 5.3.4) within one week before and within 9 days to 7 months after surgery.

Results

The mean preoperative spherical equivalent was -5.4 D and the mean postoperative equivalent was -0.3 D. The mean ablation depth was 61.7 μm. The difference of main slow corneal polarization axis was 9.8° ± 12.8° and the difference of corneal retardation magnitudes was -1.1 nm ±9.0 nm. Using the same preoperative customized corneal compensation before and after LASIK, 11 of 16 SLP parameters of the RNFL changed significantly. However, SLP parameters after LASIK with new postoperative customized corneal compensation revealed no significant difference to preoperative compensated values.

Conclusions

Our results show that there are significant changes of corneal birefringent properties induced by LASIK and that these changes can be detected and ruled out by scanning laser polarimetry with variable but not with fixed corneal compensation. Therefore, the new variable cornea compensator may be of great clinical importance in glaucoma diagnostics because measurements are less affected by the birefringent properties of anterior eye segments.

Supported by the Alfried Krupp von Bohlen und Halbach-Stiftung, Essen, Germany