gms | German Medical Science

102. Jahrestagung der DOG

Deutsche Ophthalmologische Gesellschaft e. V.

23. bis 26.09.2004, Berlin

In vitro experiments to evaluation of optimal edges of lens optics for prevention of posterior capsular opacification

Meeting Abstract

Suche in Medline nach

  • corresponding author A. Wildeck - Department of Ophthalmology, University Hospital of Berlin (Charité), Berlin; Eye Center Alt-Moabit, Berlin
  • A. Sänger - Department of Ophthalmology, University Hospital of Berlin (Charité), Berlin; Eye Center Alt-Moabit, Berlin
  • M. Tetz - Department of Ophthalmology, University Hospital of Berlin (Charité), Berlin; Eye Center Alt-Moabit, Berlin

Evidenzbasierte Medizin - Anspruch und Wirklichkeit. 102. Jahrestagung der Deutschen Ophthalmologischen Gesellschaft. Berlin, 23.-26.09.2004. Düsseldorf, Köln: German Medical Science; 2004. Doc04dogP 074

Die elektronische Version dieses Artikels ist vollständig und ist verfügbar unter: http://www.egms.de/de/meetings/dog2004/04dog565.shtml

Veröffentlicht: 22. September 2004

© 2004 Wildeck et al.
Dieser Artikel ist ein Open Access-Artikel und steht unter den Creative Commons Lizenzbedingungen (http://creativecommons.org/licenses/by-nc-nd/3.0/deed.de). Er darf vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden, vorausgesetzt dass Autor und Quelle genannt werden.


Gliederung

Text

Objective

To develop a new method for the evaluation of the optic edge design of intraocular lenses.

Methods

PMMA-lenses were specially manufactured with 11 different edge designs. Each lens design was evaluated using the programm EPCO 2000. Also its ability to stop cell growth was observed by placing each lens into cell culture and observing the cell growth over a period of 18 days on average.

Results

Out of the 11 different edge design groups only the first 3 groups with the sharpest edge design were able to prevent the growth of lens epithelial cells into the visual axis of the lens. The edge design which effectively stops cell growth is characterized by an area above the edge of 13,5μm2 at the most.

Conclusions