Objectives: Reverse Total Shoulder Arthroplasty (RTSA) is the standard of care for patients with glenohumeral osteoarthritis and rotator cuff deficiency. Pre-operative RTSA planning based on medical images and patient-specific instruments has been established over the last decade.
This study aims to determine the effects of using augmented reality assisted intraoperative navigation (ARIN) for baseplate positioning in RTSA compared to preoperative planning. It is hypothesized that ARIN will decrease deviation between preoperative planning and postoperative baseplate positioning. Moreover, ARIN will decrease deviation between the (senior) more (> 50 RTSA/year) and less experienced (junior) surgeon (5–10 RTSA/year).
Methods: Preoperative CT scans of sixteen fresh-frozen cadaveric shoulders were obtained. Baseplate placements were planned using a customized software (My Shoulder, Medacta International, Castel San Pietro, Switzerland). The data was then converted and uploaded to the augmented reality system (NextAR, Medacta International, Castel San Pietro, Switzerland). Each of the eight RTSAs were implanted by a senior and a junior surgeon, with four RTSAs utilizing ARIN and four without. The scanned scapulae were segmented, and the preoperative scan was laid over the postoperative scapula by nearest iterative point cloud analysis. The deviation from the planned entry point and trajectory was calculated regarding the inclination, retroversion, medialization (reaming depth)/lateralization, antero-posterior position and superior/inferior position of the baseplate. Data are reported as mean ± standard deviation (SD) or mean and 95% confidence interval (CI). P values <0.05 were considered statistically significant.
Results and conclusion: The use of ARIN yielded a reduction in the absolute difference between planned and obtained inclination from 9° (SD: 4°) to 3° (SD°:2) (p=0.011). Mean difference in planned-obtained inclination between surgeons was 3° in free-hand surgeries (95%CI: -4,10, p=0.578), while this difference reduced to 1° (95%CI: -6, 7, p=0.996) using ARIN.
Retroversion, medialization (reaming depth)/lateralization, antero-posterior position and superior/inferior position of the baseplate were not affected by using ARIN. Surgical duration was increased using ARIN for both the senior (10 mins) and junior (18 mins) surgeon.
The implementation of augmented reality assisted intraoperative navigation (ARIN) leads to greater accuracy of glenoid component placement, specifically with respect to inclination. Furthermore, ARIN allows less experienced surgeons to achieve a similar level of accuracy in component placement comparable to more experienced surgeons.
However, the potential advantages of ARIN in RTSA are counterbalanced by an increase in operative time.