gms | German Medical Science

64. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie (DGNC)

Deutsche Gesellschaft für Neurochirurgie (DGNC) e. V.

26. - 29. Mai 2013, Düsseldorf

First clinical experience using a portable infrared pupillometer on a neurosurgical intensive care unit

Meeting Abstract

Suche in Medline nach

  • Patrick Schiebel - Neurochirurgische Klinik, Universitätsklinikum Heidelberg
  • Martin Jakobs - Neurochirurgische Klinik, Universitätsklinikum Heidelberg
  • Andreas Unterberg - Neurochirurgische Klinik, Universitätsklinikum Heidelberg

Deutsche Gesellschaft für Neurochirurgie. 64. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie (DGNC). Düsseldorf, 26.-29.05.2013. Düsseldorf: German Medical Science GMS Publishing House; 2013. DocMO.13.09

doi: 10.3205/13dgnc114, urn:nbn:de:0183-13dgnc1144

Veröffentlicht: 21. Mai 2013

© 2013 Schiebel et al.
Dieser Artikel ist ein Open Access-Artikel und steht unter den Creative Commons Lizenzbedingungen (http://creativecommons.org/licenses/by-nc-nd/3.0/deed.de). Er darf vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden, vorausgesetzt dass Autor und Quelle genannt werden.


Gliederung

Text

Objective: Testing of the pupillary light reflex is probably the most commonly used neurological test on intensive care units when assessing unconcious patients. Although of utmost clinical importance, testing of the pupillary reflex has always been susceptible to subjective judgement of the investigator and performed using non-standardized penlights. In our age of high-tech medicine means to quantify and objectify the pupillary reflex seem to be a logical development.

Method: We tested the portable infrared monocular pupillometer NPi™-100 by NeurOptics® in 47 patients on our neurosurgical ICU. The pupillometer measures pupil size, constriction and dilation velocity, constriction latency of which it calculates the Neurological Pupil index (NPi). The NPi ranges between 0 and 5: Values between 4 and 5 indicate normal, values between 0 and 3 pathological pupillary function. Measurements where made taking consideration of brain impairment (assessed with radiological or electroencephalic data), intracerebral pressure (ICP), narcotics and conventional pupillary testing.

Results: In isochoruous patients the NPi-values were normal even with severe brain impairment. When anisocoria was present NPi-values in general were lower ranging between 3 and 0. We observed deterioration of NPi-values of patients with anisocoria concomittant to clinical worsening. Both false-positive and false-negative results for conventional testing of pupillary reactivity compared to the pupillometer could be observed. The pupillometer was able to detect a pupillary reaction in patients with pronounced myosis due to high-dose opioid sedation when conventional testing did not show any reactivity at all.

Conclusions: The portable infrared NPiTM-100 pupilometer is a useful resource-friendly tool to objectify commonly aquired data on a neurosurgical ICU. However further and systematic investigations have to be performed in the future to correlate NPi-values to other clinical parameters like ICP or PtiO2 and to radiographic changes in pathology. Development of binocular measurement of direct and indirect pupil reactivity seems to be useful in neurosurgical patients to establish infrared pupillometry as a standard procedure.