Primary stability of open wedge high tibial osteotomy: a biomechanical study of four implants
Suche in Medline nach
Autoren
Veröffentlicht: | 13. Juni 2005 |
---|
Gliederung
Text
Introduction
This biomechanical study investigated the primary stability of different implants for open wedge high tibial osteotomy (HTO).
Materials and methods
17 sawbone 3rd generation composite tibiae were used for testing. 4 plates with different designs were tested: short spacer plate (OWO) (n=4), short spacer plate with multidirectional locking bolts (MSO) (n=5), prototype version of long spacer plate with multidirectional locking bolts (MSOnew), (n=2), long medial tibia plate fixator (MPF) (n=4). All osteotomies were performed by one surgeon (PL) in a standardized fashion. The proximal and distal end of the tibia were embedded in a specially designed mounting fixture. Axial compression of the tibiae with a standardized alignment of the loading axis was performed using a material testing machine. Single load to failure tests as well as load-controlled cyclical failure tests were performed. The required force and time to failure were recorded as well as the gap motion between the proximal and distal segment using linear displacement transducers.
Results
Failure occurred at the lateral cortex in all tested implants. Significant differences were found regarding force and time to failure with the MPF being the construct with the highest resistance. Displacements at the important medial osteotomy gap were significantly different. The MPF showed the lowest medial displacement, followed by the OWO, MSO and at least by the MSOnew plate.
Conclusion
The implant design strongly influences the primary stability of open wedge HTO. The best results were obtained with a long plate fixator with locked bolts.