gms | German Medical Science

55. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie e. V. (DGNC)
1. Joint Meeting mit der Ungarischen Gesellschaft für Neurochirurgie

Deutsche Gesellschaft für Neurochirurgie (DGNC) e. V.

25. bis 28.04.2004, Köln

Identification of cerebral aneurysm clips to determine MRI-compatibility using the “ClipFinder”

Identifikation zerebraler Aneurysmenclips zur Beurteilung der MR-Kompatibilität mit Hilfe des "ClipFinder"

Meeting Abstract

  • corresponding author Karsten Schöller - Neurochirurgische Klinik, Klinikum Großhadern, Ludwig-Maximilians-Universität, München
  • D. Morhard - Abteilung für Neuroradiologie, Klinikum Großhadern, Ludwig-Maximilians-Universität, München
  • H.-J. Steiger - Neurochirurgische Klinik, Heinrich-Heine-Universität, Düsseldorf
  • R. Schmid-Elsaesser - Neurochirurgische Klinik, Klinikum Großhadern, Ludwig-Maximilians-Universität, München

Deutsche Gesellschaft für Neurochirurgie. Ungarische Gesellschaft für Neurochirurgie. 55. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie e.V. (DGNC), 1. Joint Meeting mit der Ungarischen Gesellschaft für Neurochirurgie. Köln, 25.-28.04.2004. Düsseldorf, Köln: German Medical Science; 2004. DocDI.08.02

Die elektronische Version dieses Artikels ist vollständig und ist verfügbar unter: http://www.egms.de/de/meetings/dgnc2004/04dgnc0224.shtml

Veröffentlicht: 23. April 2004

© 2004 Schöller et al.
Dieser Artikel ist ein Open Access-Artikel und steht unter den Creative Commons Lizenzbedingungen (http://creativecommons.org/licenses/by-nc-nd/3.0/deed.de). Er darf vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden, vorausgesetzt dass Autor und Quelle genannt werden.


Gliederung

Text

Objective

Knowledge of the magnetic properties of cerebral aneurysm clips in patients undergoing magnetic resonance imaging (MRI) is imperative. However in daily practice, it usually is impossible to derive this information from earlier medical records. The aim of this study was to develop a free accessible internet database to identify unknown cerebral aneurysm clips based on conventional X-ray images of the skull, and to provide reliable information about their magnetic properties to allow safe MRI-investigations in these patients.

Methods

The 17 most commonly used clip-series from the last 50 years were selected. The ferromagnetic properties of 47 example clips were examined by the FDA-approved “deflection angle test”. A HTML-based database was developed using standard X-ray images in two planes and photographs of the clips. Furthermore, details about alloy, ferromagnetic properties, MRI-compatibility, year of construction, and other characteristics identifiable on x-rays were included.

Results

Seven aneurysm clips (of 4 clip-models) were judged to be not MRI-compatible. The database named “ClipFinder” allows easy identification of cerebral aneurysm clips on standard skull x-rays by comparing the films with the databank images supported by information on unique features of the clip design. Preliminary testing of 18 clips on standard X-ray and DSA-images by an investigator, blinded to relevant clip-data and patient history, resulted an accuracy of 100% for MRI-compatibility and of 85% for exact identification of all implanted clips.

Conclusions

“ClipFinder” provides a fast, inexpensive, and easy to use database to identify unknown cerebral aneurysm clips via the internet, using standard x-ray images. Information on the clips’ ferromagnetic properties allows safe MRI studies.