Marker free referencing for neuronavigation: A new technique
Markerlose Referenzierung für die Neuronavigation: neue Technik für Patientenerfassung und -Tracking
Search Medline for
Authors
Published: | April 23, 2004 |
---|
Outline
Text
Objective
There are several possibilities for referencing of a neuronavigation system. Until now pointer based referencing using preoperative implanted bone screws is golden standard for this procedure. Nevertheless skin fiducials worldwide are used for the majority of navigated interventions. There have been several attempts to overcome the intrinsic drawbacks of the use of skin fiducials. We present a new, marker free referencing system for neuronavigation without the need for rigid head fixation.
Methods
We used the Stryker Navigation System and the Face-Mask-Referencing and Tracking device in a cadaver study. The accuracy of face mask referencing and tracking was compared to bone screw based referencing with mayfield fixation. 50 measurements of accuracy during 5 referencing procedures of each of these systems were performed using 8 control markers (bone screw markers) implanted on the cadaver head.
Results
There is no need for additional imaging for neuronavigation as the system needs no markers either on the patient or on the data set. The face mask referencing and tracking device consists of a self adhesive single use foil with 31 LED fixed on it. After the mask is fixed to the patients face, the Stryker navigation system is able to do an automatic referencing to the calculated surface of the patient's 3-D-Dataset. Mayfield fixation is not necessary as the face mask can be used as a tracking device. The accuracy that was achievable with the face mask was 1.1 mm compared to 1.2 mm of bone screw referencing.
Conclusions
The Face-Mask-Referencing and Tracking device is easy to use and facilitates the preoperative preparation time by automatic referencing without fiducials on patient and image data set and without the need for rigid head fixation. Using our new device we could achieve an accuracy comparable to that of bone screw markers. We already used the system for ENT-Interventions and transsphenoidal pituitary gland surgery.