gms | German Medical Science

76. Jahresversammlung der Deutschen Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf- und Hals-Chirurgie e. V.

Deutsche Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf- und Hals-Chirurgie e. V.

04.05. - 08.05.2005, Erfurt

Experimental assessment of electrode insertion forces – a training tool to control surgical skills

Meeting Abstract

Suche in Medline nach

  • corresponding author Dirk Mürbe - Dept. of Otorhinolaryngology, University of Technology, Dresden, Germany
  • Matthias Bornitz - Dept. of Otorhinolaryngology, University of Technology, Dresden, Germany
  • Thomas Zahnert - Dept. of Otorhinolaryngology, University of Technology, Dresden, Germany

Deutsche Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf- und Hals-Chirurgie. 76. Jahresversammlung der Deutschen Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf- und Hals-Chirurgie e.V.. Erfurt, 04.-08.05.2005. Düsseldorf, Köln: German Medical Science; 2005. Doc05hno561

Die elektronische Version dieses Artikels ist vollständig und ist verfügbar unter: http://www.egms.de/de/meetings/hno2005/05hno026.shtml

Veröffentlicht: 22. September 2005

© 2005 Mürbe et al.
Dieser Artikel ist ein Open Access-Artikel und steht unter den Creative Commons Lizenzbedingungen (http://creativecommons.org/licenses/by-nc-nd/3.0/deed.de). Er darf vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden, vorausgesetzt dass Autor und Quelle genannt werden.


Gliederung

Text

Cochlear implantation surgery aims to prevent cochlear structures by minimization of insertional trauma. However, intraoperatively, the feedback of electrode insertion is limited to the manual assessment of the implanting surgeon. The aim of the present study was to analyse electrode insertion by measurement of applied forces in dependence of different kinds of insertion failure and microsurgical experience.

Six otosurgeons of varying microsurgical experience were asked to insert practice electrodes in a cochlear model. Insertion forces were measured by means of a force cell connected with an electrode claw and digitally stored.

Maximum and mean force values were found to depend on the subject’s microsurgical experience. Insertion failures like tip fold were related to changed insertion forces.

Force measurements during electrode insertion provide additional feedback to the implanting surgeon. This feedback might support the training of surgical skills for an improved performance of electrode insertion in cochlear implantation.