Visualisation of the anterior cerebral artery complex with a continuously variable-view rigid endoscope: An anatomical study

Meeting Abstract

  • Florian H. Ebner - Klinik für Neurochirurgie, Eberhard-Karls-Universität, Tübingen, Germany
  • Martin U. Schuhmann - Klinik für Neurochirurgie, Eberhard-Karls-Universität, Tübingen, Germany
  • Jakob S. Marquardt - Institut für Klinische Anatomie, Eberhard-Karls-Universität, Tübingen, Germany
  • Bernhard Hirt - Institut für Klinische Anatomie, Eberhard-Karls-Universität, Tübingen, Germany
  • Marcos Tatagiba - Klinik für Neurochirurgie, Eberhard-Karls-Universität, Tübingen, Germany

Deutsche Gesellschaft für Neurochirurgie. 61. Jahrestagung der Deutschen Gesellschaft für Neurochirurgie (DGNC) im Rahmen der Neurowoche 2010. Mannheim, 21.-25.09.2010. Düsseldorf: German Medical Science GMS Publishing House; 2010. DocP1744

doi: 10.3205/10dgnc215, urn:nbn:de:0183-10dgnc2155

Veröffentlicht: 16. September 2010

© 2010 Ebner et al.
Dieser Artikel ist ein Open Access-Artikel und steht unter den Creative Commons Lizenzbedingungen (http://creativecommons.org/licenses/by-nc-nd/3.0/deed.de). Er darf vervielfältigt, verbreitet und öffentlich zugänglich gemacht werden, vorausgesetzt dass Autor und Quelle genannt werden.


Gliederung

Text

Objective: Neuroendoscopy is increasingly used as an adjunctive tool in intracranial aneurysm surgery. The aim of the present anatomical study was to assess the versatility of a prototype continuously variable-view rigid endoscope in visualising the anterior cerebral artery complex.

Methods: In five formaldehyde fixed, arterially injected specimens a standard frontolateral approach was performed on both sides. After meticulous microsurgical dissection of the appraoch, the prototype of a multivariable rigid endoscope (EndoCAMeleon, Karl Storz, Tuttlingen, Germany) was inserted for inspecting the anterior cerebral artery complex, utilising and testing the facilities of the device.

Results: The continuously variable viewing mechanism enables the surgeon to adjust the field of view continuously and to optimise the visualisation of the neurovascular structures. Due to the rigid tip combined with the continuously variable viewing mechanism, the necessity to move the endoscope within the surgical field was minimal. The field of view changes but the tip itself hardly moves. The prototype was able to enhance both the visibility of the anterior cerebral artery complex as well as the accessibility of the A1 and A2 arterial walls to a range of approximately 270°.

Conclusions: The endoscope with continuous variable viewing mechanism enhances the visibility of the anterior cerebral artery complex and facilitates endoscope-assisted inspection, planning of clip application and control of clip position. The device has the potential to become a most useful, safe and easy to use adjuvant tool in aneurysm surgery.